    摘要:

    原位制备了羧甲基壳聚糖/铜复合物修饰电极、羧甲基壳聚糖/银复合物修饰电极和壳聚糖/聚苯胺/钯复合物修饰电极,金属粒子均匀分散于聚合物薄膜中.3种修饰电极对H2 O2 均表现出了良好的检测性能,它们对H2 O2 的线性响应范围依次为1.0 μmol/L~10.0 mmol/L, 10 μmol/L~6.0 mmol/L和10 μmol/L~7.0 mmol/L,相关系数依次为0.999 7, 0.999 9和0.999 8,灵敏度依次为33 μA·L·mmol -1 ·cm -2 ,37 μA·L·mmol -1 ·cm -2 和46 μA·L·mmol -1 ·cm -2 ,检测限依次为:1 μmol/L,1 nmol/L和1 μmol/L.

    Abstract:

    Carboxymethyl chitosan/Cu composite, carboxymethyl chitosan/Ag composite and chitosan/polyaniline/Pd composite modified electrodes were prepared through in situ electrodeposition. The resultant metal particles were uniformly dispersed in the polymer films. The three electrodes showed good analytic performances to H2O2 with the linear H2O2 concentration ranges of1.0mol/L~10.0mmol/L,10mol/L~6.0mmol/L,10mol /L~7.0 mmol/L, the related coefficients corresponding to 0.999 7, 0.999 9 and 0.999 8, the sensitivities corresponding to 33, 37 and 46 μA·L·mmol -1 ·cm -2 , and the detection limits corresponding to 1 μmol/L, 1 nmol/L and 1 μmol/L, respectively.